- BEMC
- BEMC Detector Operator Manual
- Calibrations
- BPRS
- BSMD
- 1) DATA: 2008 BSMD Calibration
- 01) raw spectra
- 02) relative BSMD-E gains from 1M dAu events
- 03) more details , answering Will
- 04) bad CAP 123
- 05) BSMDE saturation, dAu, 500K minB eve
- 06) QAed relative gains BSMDE, 3M d-Au events , ver1.0
- 07) QA method for SMD-E, slopes , ver1.1
- 08) SMD-E gain equalization , ver 1.1
- 09) QA of SMD-P slopes, ver1.1
- 10) SMD-P gain equalization , ver 1.1
- 12) investigating status of P-strips
- 13) ver 1.2 : SMD-E, -P, status & relative gains, no Crate4
- 14 Eval of BSMDE status tables for pp 2008, day 49,50
- 15 stability of BSDM peds, day 47 is good
- 15a ped stability day 47, take 2
- 16) Time stability by fill of BSMD pedestals
- 17) Absolute gains , take1
- 18 Absolute gains, take 2
- 19) Absolute BSMD Calibration, table ver2.0, Isolated Gamma Algo description
- 20 BSMD saturation
- 1) M-C : response of BSMD , single particles (Jan)
- details about known hardware problems
- details of SMD simulator, simu shower zoom-in
- Mapping, strip to tower distance
- one cluster topology , definition of 'barrel cell'
- 1) DATA: 2008 BSMD Calibration
- Run 3 BTOW Calibration
- Run 4 BTOW Calibration
- Run 5 BTOW Calibration
- Run 6 BTOW Calibration
- Run 7 BTOW Calibration
- Run 8 BTOW Calibration (2008)
- Database
- Mapping
- Service Tasks
- Software
- Useful Documents
- DAQ
- Detector Upgrades
- EEMC
- FTPC
- HFT
- L3
- PMD
- RICH
- Slow Controls
- SSD
- SVT
- TOF
- TPC
- Trigger Detectors
15a ped stability day 47, take 2
Submitted by balewski on Sun, 2008-08-10 16:03.
Under:
On August 8, BSMD peds in the offline DB where corrected for day 47.
Runs minb 34 & 74 were used to determine and upload DB peds.
Below I evaluated pedestal residua for 2 runs : 37 & 70, both belonging to the same RHIC fill.
I have used 500 zero-bias events from runs 37 & 70, for the official production w/o zero suppression.
All strips for which mTables->getStatus(iEP, id, statPed,"pedestal"); returns !=1 and all events using CAP123,124,125 were dropped.
Fig 1,2 show big picture: all 38,000 strips.
Fig 3 is zoom in on some small & big problems.
Fig 4 & 5 illustrates improvement if run-by-run pedestal is used.
Fig 1, run=9047037

Fig 2, run=9047070

Fig 3, run=9047070, zoom in

Private peds were determine for 16 runs for day 47 and used appropriately. Below is sum of ped residua from all 16 runs, from zero bias events.
Fig 4, run=9047001,...,83 zoom in

Fig 5, run=9047001,...,83 full range