If MiniDAQ fails to get events from the readout boards, this is most
likely to be a problem with the electronics setup or cabling. Common
problems include:
- RDO's not powered:
FEE power supplies
must be on and properly connected for all enabled readout boards.
- Wrong Rosie-busy configuration: Verify that the
synchronization module
setup is correct, with Rosie busies enabled (switch up) if, and
only if, the Rosie was enabled at configuration time.
- No triggers:
- The LeCroy 222 gate/delay NIM module for the trigger output
has a red LED that flashes visibly for each trigger; if it
never flashes, check the trigger source and enable in the
trigger crate.
- The RDO's have a green LED that flashes visibly for each
trigger; if no readout board flashes, check the
trigger card
and the
trigger cable.
See also: Data comes slowly