EPD Gate Timing Scan - Run 2024

The procedure for this exercise has been described before.  In 2018, Mike wrote up the approach and code and later Joey wrote up a nice set instructions.

We have eight "labels" in the trigger configuration, allowing us to set separately the gates of the ADCs.  My recommended values for the gate START are in red.
  • East QTBs for EQ 1,2,3 - 60
  • East QTCs for EQ 1,2,3 - 55
  • West QTBs for EQ 1,2,3 - 75
  • West QTCs for EQ 1,2,3 - 75
  • East QTBs for EQ 4 - 72
  • East QTCs for EQ 4 - 75
  • West QTBs for EQ 4 - 75
  • West QTCs for EQ 4 - 85

The other label values are directly derived from these START values:
  • QTB_STOP = QTB_START - 11
  • QTC_STOP = QTC_START - 18
  • TAC_STOP = 2*START - 20

The proper timing labels for the EPD are as follows  (I have not completed filling in the following table yet.)

EPD-TACdiff-Max 4095
EPD-TACdiff-Min 50
EPD_EQ4_QTb_E_Gate_End_Delay  
EPD_EQ4_QTb_E_Gate_Start_Delay 72
EPD_EQ4_QTb_W_Gate_End_Delay  
EPD_EQ4_QTb_W_Gate_Start_Delay 75
EPD_EQ4_QTc_E_Gate_End_Delay  
EPD_EQ4_QTc_E_Gate_Start_Delay 75
EPD_EQ4_QTc_E_TAC_Stop  
EPD_EQ4_QTc_W_Gate_End_Delay  
EPD_EQ4_QTc_W_Gate_Start_Delay 85
EPD_EQ4_QTc_W_TAC_Stop  
EPD_QTb_E_Gate_End_Delay  
EPD_QTb_E_Gate_Start_Delay 60
EPD_QTb_W_Gate_End_Delay  
EPD_QTb_W_Gate_Start_Delay 75
EPD_QTc_E_Gate_End_Delay  
EPD_QTc_E_Gate_Start_Delay 55
EPD_QTc_E_TAC_Stop  
EPD_QTc_W_Gate_End_Delay  
EPD_QTc_W_Gate_Start_Delay 75
EPD_QTc_W_TAC_Stop  
   
   
   


I'll probably put some figures or t




For prepost=0, you can find the full gate scan here.  And here is the one for prepost=1.  To be honest, I would like to talk to the trigger folks about this, as it seems you can set the gate delays to different values and pick up the signal in a different prepost.  How close together are these preposts, anyway?