dE/dx calibration of Y2001 data (progress report)

  1. Data sample
  2. dE/dx resolution after pulser corrections only.
  3. Time dependence (done)
  4. Sector and Row dependence (almost done)
  5. Sector and Z dependence (in progress)
  6. pile-up and volume charge (not yet)

Data sample

dE/dx resolution after pulser corrections only.

Resolution at 90cm = 8.7%.

Time dependence

Overall scale is about exp(-0.1). There are 3 runs with very high scale (scale > 0.5) and many runs with scale < -0.2.
The result for these runs have to understood and the runs have to excluded from analysis.

Sector and Row dependence

After the pulser correction (pad level) still exits dependence of Sector and Row: (Z limits are set -0.5/+0.5)

It is needed one more iteration.


Sector and Z dependence

We are looking for effects due electron absorption.

Result for inner and outter sector

Dependence on no. of global tracks(pile-up and volume charge)

Yuri Fisyak
Last modified: Thu Dec 6 12:51:56 EST 2001