Tower (red) trouble from "Read All/Reload":


  1. One can try and repeat the fill start "Master Reload" from the tower FEE GUI. This is the easiest approach; Master Reload does a reset and reconfiguration of all crates (along with calculation and redownload of all LUT's). This complete operation takes ~ 5 minutes including the final crate checking sequence (sometimes required to get final ready status).

  2. But more directly (or if "Master Reload" has [repeated*] difficulty for some tower crates), action at the lower level individual screens of the affected crates can be employed (to access, click on the corresponding gray box labeled "Crate #n" from the tower FEE GUI). A "load", "write", "read" writes and reads the configuration and one should see the individual FEE cards go from COMMAND to RUN mode in the final phase. Note: the LUTs are calculated/loaded automatically during the next run start so "calc LUT" and "write LUT" are not explicitly needed.

    • a hard "reset" (popup window on "resets" from individual tower crate GUI) reprograms all FPGA's to power up value and hence must be followed by complete reconfiguration (LUTs are downloaded upon run start) as per a power cycle.

    If a power cycle is employed to clear problems, or if any tower FEE crates have tripped off, a "Master Reload" (or its equivalent as described above at the crate level - recall there are 2 crates per supply!) is required after resetting the LV power.

  3. When in doubt, consult an EEMC expert!


    *[Note: the "break" button may be used to interrupt master reload, thus allowing corrective actions to be taken; at the end, a final full reload will of course be required.]