Acceptance of the NA44 spectrometer is discussed in section 3.7
and presented on Fig. 3.3.
The
spectra fall with
and their shapes within the NA44 acceptance
can be justifiably fit [93] by a single
.
The will be referred to as a temperature parameter
or a slope parameter.
In determining and
for kaons and pions we use spectrometer
settings, or portions thereof, with
.
The slope parameters do not depend appreciably on within this small range.
Then the double differential multiplicity can be factorized
as
(10)
In parallel to the equations (4.7),
for the observable particle counts one can define
(11)
(12)
(Technically speaking, they are projections of 2D histograms.)
Integration of Eq. 4.5 assuming that
Eq. 4.8 holds yields
(13)
If Eq. 4.8 is true, from Eq. 4.7
can be rewritten as
(14)
Eliminating , we conclude the ``theoretical justification''
for what is known as a 1D acceptance correction:
(15)
Similarly, for the spectrum we get
(16)
Similar equations can be obtained for or .
In principle, one needs to know and in order to determine
the true shape of spectrum and .
In practice, for pions was taken to be a Gaussian with
centered at mid-rapidity; for kaons - a Gaussian with
.
For the acceptance correction in , I used
(17)
Here are 8 variable parameters, are fixed ``temperatures''.
For pions, all 8 parameters
were used (see Table 4.1);
for kaons, good fits could be obtained with only one parameter
(see Table 5.1).
The original idea of fitting the pion distributions with the
multi-temperature formula 4.15
to test sufficiency of a simple statistical description with a single
temperature and a chemical potential,
by transforming the spectrum into a ``temperature spectrum''.
The pion spectra seem to be dominated by a temperature component around 200 MeV.
Table 4.1:
Parameters of the multi-temperature fits to the transverse kinetic
energy distributions of pions.
See Section 4.1.2, Eq. 4.15.
The fitting range is